Type:
Journal
Description:
A conventional scanning electron microscope equipped with a LaB6 source has been modified to operate in a scanning transmission mode. Two detection strategies have been considered, one based on the direct collection of transmitted electrons, the other on the collection of secondary electrons resulting from the conversion of the transmitted ones. Two types of specimens have been mainly investigated: semiconductor multilayers and dopant profiles in As-implanted Si. The results show that the contrast obeys the rules of mass–thickness contrast whereas the resolution is always defined by the probe size independently of specimen thickness and beam broadening. The detection strategy may affect the bright field~ light regions look brighter! or dark field~ heavy regions look brighter! appearance of the image. Using a direct collection of the transmitted electrons, the contrast can be deduced from the angular distribution of transmitted electrons and their collection angles. When collecting the secondary electrons to explain the image contrast, it is also necessary to take into account the secondary yield dependence on the incidence angle of the transmitted electrons.
Publisher:
Cambridge University Press
Publication date:
1 Feb 2005
Biblio References:
Volume: 11 Issue: 1 Pages: 97
Origin: